TECNICAL CHARACTERISTICS
Type II Water (Analytical Grade): The most demanded water in Hospitals and Clinical analysis laboratories
Autwomatic CLINICAL 200-400 model produce from 200 to 400 l/h Type II Purified Water from tap water
Versions |
Liter/hour |
Purification Module |
UV Lamp |
Final filter 0.22 micron |
By Pass Security |
AUTWOMATIC CLINICAL 200 |
200 |
X |
X |
X |
X |
AUTWOMATIC CLINICAL 300 |
300 |
X |
X |
X |
X |
AUTWOMATIC CLINIAL 400 |
400 |
X |
X |
X |
X |
Specially designed to automatically feed to clinical analyzers.
Reduces the number of stops for maintenance and change of consumables.
Systems Pressure:Type II Water newly produced, without storage in a final tank
The water supplied to Autoanalyzer is newly produced.Differently from other equipment, there is no storage of the final water, in order to avoid its deterioration.The Osmotic Water is stored in pressure tanks of 50, 100 and 250 liters.
Atmospheric Systems
When the customer requires, our systems can work with atmospheric deposits 50,100,200 and 500 liters.
Monitoring. Continuous monitoring of the purification process parameters:
- Real number of working hours of the equipment.
- Feed water conductivity (± 1 mS/cm).
- Osmotic Water conductivity (± 1 mS/cm)
- % performance of the reverse osmosis module (± 0,1%).
- Conductivity of the produced Type II water (± 0,1 mS/cm).
- Water temperature. All conductivity measurements are compensated at 25 ºC.
- Status of the equipment (producing/full tank/dispensing).
Automatic operation
The equipment starts and stops automatically, thus guarantying the production of Type II Water. Automatic stop in case of water supply cut.
Parameters customizing
The user can fix different parameter values, according to the control requirements:
- Pre-treatment time
- Minimum performance of the reverse osmosis module.
- Maximum conductivity of the produced Type II water.
Warnings
The equipment warns, by a visual and audio message, of the need to replace the consumables, according to the fixed values.
- Pre-treatment change
- Interchange resin change
- Change of the reverse osmosis module
- Warning of malfunction of conductivity and temperature probe
Ultraviolet lamp and Final Filter
The disinfection equipment emits ultraviolet radiation in two wave lengths, leading to a reduction of the bacterial content of 5 logaritmic cycles. This provides the user with water with minimum bacterial contamination if needed and produces the oxidation of organic material traces in the water.
For those laboratory applications where the bacterial contamination is critical, a final filter of 0,22 microns ensures a bacterial count ≤ 1 ufc/ml.
Security system via By-pass
When Autwomatic is supplying autoanalyzers, Wasserlab has designed an auxiliar system able to produce Type II Water directly from tap water, without generation of burbles, reducing the stops of autoanalizers.
Security system by double sensor monitorization
The equipment includes two conductivity sensors of Type II Water. This allows to detect that first deionization module is exhausted. On this way it allows to plan maintenance operations on time and assuring the continuous work of autoanalyzer.