Autwomatic RO Process 200-400

TECNICAL CHARACTERISTICS

Osmotized water

Autwomatic RO High production models produce from 200 to 400 l/h Type III water from tap water.

Versions

Liter/hour

Purification Module

UV Lamp

Final Filter 0.22 micron

By Pass Security

AUTWOMATIC RO PROCCESS 200

200

-

-

-

-

AUTWOMATIC RO PROCCESS 200 UV

200

-

X

X

-

AUTWOMATIC RO PROCCESS 300

300

-

-

-

-

AUTWOMATIC RO PROCCESS 300 UV

300

-

X

X

-

AUTWOMATIC RO PROCCESS 400

400

-

-

-

-

AUTWOMATIC RO PROCCESS 400 UV

400

-

X

X

-

Based on a sequential purification steps :sediment filtration, dechlorination, reverse osmosis, and optionally Ultraviolet treatment and final filtration of 0.22 micron.

Ideal to supply water to industrial processes and machines requiring purified water with the lower cost per liter.

Monitoring

Continuous monitoring of the purification process parameters:

-   Real number of working hours of the equipment.
-   Feed water conductivity (± 1 mS/cm).
-   Osmotic water conductivity (± 1 mS/cm)
-   % performance of the reverse osmosis module (± 0,1%).
-   Water temperature. All conductivity measurements are compensated at 25ºC.
-   Status of the equipment (producing/full tank/dispensing).

 


Automatic operation

The equipment starts and stops automatically, thus guarantying the production of Osmotized  water.

Automatic stop in case of water supply cut.

Parameters customizing

The user can fix different parameter values, according to the control requirements:

-      Pre-treatment time.

-      Minimum performance of the reverse osmosis module.

Warnings

The equipment warns, by a visual and audio message, of the need to replace the consumables, according to the fixed values:

-      Pretreatment change.

-      Interchange resin change.

-      Change of the reverse osmosis module.

-      Warning of malfunction of conductivity and temperature probes.

 

The equipment warns by a visual ans audio message of the need to replace consumables, according to the fixed values.

Pretreatment reinforcement

In case of feed water containing high levels of solids or high hardness, it is convinient to reinforce the Standard pretreatment with:

 

-        Mineral FIlter, automatic, with autocleaning device. 

-        Softener: Reduces water hardness, increasing the live of equipment.

Ultraviolet lamp

The disinfection equipment emits ultraviolet radiation in two wave lengths, leading to a reduction of the bacterial content of 5 logaritmic cycles. This provides the user with water with minimum bacterial contamination if needed and produces the oxidation of organic material traces in the water.

Open tank and Pressurized tank

The equipment can be supplied with an pressurized tank where water is stored avoiding the contact with air or Light.

 

When high storage volume are necessary, open tanks are supplied, with electrical level detection and overflow. 

Conduct meter calibration and system verification

Autwomatic software allows the adjustment and calibration of the conductivity meter by a traceable certified standard, according to the national standards of the German Deutscher Kalibrierdienst (DKD).

 

 

Hydraulic diagram