Selecciona Idioma Contact (+34) 948 186 141 info@wasserlab.com

Autwomatic Process 200-400

TECNICAL CHARACTERISTICS

Type II Water (Analytical Grade): Purified Water supply for machines and industrial processes

Autwomatic Process produces from 200 to 400 l/h Type II Purified Water from tap water.

Versions

Liter/hour

Purification Module

UV Lamp

Final filter 0.22 micron

By Pass Security

AUTWOMATIC PROCESS 200

200

X

-

-

X

AUTWOMATIC PROCESS 200 UV

200

X

X

X

X

AUTWOMATIC PROCESS 300

300

X

-

-

X

AUTWOMATIC PROCESS 300 UV

300

X

X

X

X

AUTWOMATIC PROCESS 400

400

X

-

-

X

AUTWOMATIC PROCESS 400 UV

400

X

X

X

X

Based on a sequential purification steps (sediment filtration, dechlorination, reverse osmosis, deionization, Ultraviolet treatment and final filtration of 0.22 micron.

Ideal to supply water to industrial processes and machines requiring purified water with the lower cost per liter.

Reduces drastically the number of stops to make changes of resin. Easy, simple and fast maintenance.

Monitoring

Continuous monitoring of the purification process parameters:

-   Real number of working hours of the equipment.
-   Feed water conductivity (± 1 mS/cm).
-   Osmotic water conductivity (± 1 mS/cm)
-   % performance of the reverse osmosis module (± 0,1%).
-   Conductivity of the produced Type II Water (± 0,1 mS/cm).
-   Water temperature. All conductivity measurements are compensated at 25ºC.
-   Status of the equipment (producing/full tank/dispensing).

Automatic operation

The equipment starts and stops automatically, thus guarantying the production of Type II Water. Automatic stop in case of water supply cut.

 

Parameters customizing

The user can fix different parameter values, according to the control requirements:

- Pre-treatment time
- Minimum performance of the reverse osmosis module.

Warnings

The equipment warns, by a visual and audio message, of the need to replace the consumables, according to the fixed values.

-  Pre-treatment change
-  Change of the reverse osmosis module
-  Warning of malfunction of conductivity and temperature probes

The equipment warns by a visual ans audio message of the need to replace consumables, according to the fixed values.

Pretreatment reinforcement

When the water supply to the Equipment contains dirt or high hardness, it is recommended to reinforce the standard pretreatment with:
- Mineral Filter, automatic with self-cleaning system.
- Softener: Removes water hardness, prolonging equipment life

Ultraviolet Lamp (UV Version)

The disinfection equipment emits ultraviolet radiation in two wave lengths, leading to a reduction of the bacterial content of 5 logaritmic cycles. This provides the user with water with minimum bacterial contamination if needed and produces the oxidation of organic material traces in the water.

Security system via By-pass

System that allows a production of Type II Water directly from tap water in case of equipment is stopped.

Security system by double sensor monitorization

It is possible to use two conductivity sensors of Type II Water. This allows to detect that first deionization module is exhausted. On this way it allows to plan maintenance operations on time and assuring the continuous work of processes.

Pressure and Atmospheric Tank 

The device can be supplied with pressurized tank, where the water is stored in a sealed system protected from light and contact with air.

For large storage, atmospheric deposits of different volumes are used, with electric float and safety overflow.

Conductivity calibration and verification system

Autwomatic software allows adjustment and calibration of conductivity using certified standard traceable to national standards of the Deutscher Kalibrierdienst (DKD) German.

Hydraulic diagram



Technical Specifications

Download documents

Back to Top